Jeio Tech AAHKB058K JMV-100 Heating & Cooling Chamber 380V, 50/60Hz, 3P
Jeio Tech AAHKB058K JMV-100 Heating & Cooling Chamber 380V, 50/60Hz, 3P
Features
Performance
- Suitable for specific use such as heat treatment and drying process.
- Microprocessor PID control / Temperature calibration / Automatic tuning.
- Heat-resistant silicone packing completely blocks high heat leakage from the inner chamber.
- Two-point door latch also enhances tightness to door closure.
- Energy efficiency is achieved thanks to optimization of the insulation system.
- Even if the unit is operating at max. temperature, the surface of the unit doesn’t allow high temperatures of over 51℃. (DIN EN 563) - Reliable design and production of door is guaranteed.
- Evaluates the results after opening the door by 90 degrees and applying 50kg load to the point of 30cm from hinges.
Safety
- User selectable operation mode after power interruption
- ether Automatic run or Manual run. - Over-temperature protection.
- Over-current and current leakage protection. (ELCB, earth leakage circuit breaker)
- Independent safety fuse.
- Important parts such as controller and heaters are highly secured from short circuit. Especially each heater has an individual fuse to completely break a short circuit. - Reversed-phase protection.
- Electrical connection error indicator helps user to rapidly fix the connection. - Over-current protection of motor.
- Automatic shut-off in situations where there is over-current to the motor. - Open door warning.
- Heater and fan stop when the door is the open state. - Electric wiring panel access warning.
- Automatic stop if the cover of electric wiring panel is opened. - Built in door lock with a key.
Convenience
- Corrosion resistance stainless steel inner chamber.
- Space efficient and easy loading chamber design.
- Included two shelves for loading of large amounts of samples.
- Heavy load shelving, rated for 40kg are also available. (optional) - User friendly LCD controller.
- Colorful interactive touch type LCD controller.
- Set point temperature and present temperature are clearly shown together on display.
- Sufficient programmable control support.
- Operating progress can be shown in graph mode for better data understanding.
- Automatic start/stop can be programmed for unattended operations.
- Display indicates abnormal temperature changes or power failures.
- Three different language support. (English / Chinese / Korean)
- Temperature and other operating parameters can be recorded using a six channel digital recorder. (optional) - Computer interface.
- RS-485 communication port (optional) enables connection up to maximum 32 units. Simultaneous monitoring and operating of multiple units is an available feature.
- Program patterns are simple to set with the use of our software.
- Easy to check status of real-time operations. Graph data supports easy understanding.
- Storage and documentation of programs.
- Data with saving capabilities in excel and text format.
- Storage of program information, backup of value settings, recovery, and temperature values.
- RS-232 communication port as standard offer
Specifications
| Model | JMV-025 | JMV-040 | JMV-070 | JMV-100 |
| Temperature data | ||||
| Range (℃ / ℉) | -35 to 180 | -35 to 180 | -35 to 180 | -35 to 180 |
| / -31 to 356 | / -31 to 356 | / -31 to 356 | / -31 to 356 | |
| Fluctuation (±℃ / ±℉) 1), 2) | ||||
| at -40℃ | 0.4 / 0.72 | 0.2 / 0.36 | 0.2 / 0.36 | 0.2 / 0.36 |
| at -25℃ | 0.4 / 0.72 | 0.3 / 0.54 | 0.2 / 0.36 | 0.2 / 0.54 |
| at -10℃ | 0.5 / 0.9 | 0.2 / 0.36 | 0.2 / 0.36 | 0.2 / 0.36 |
| at 25℃ | 0.3 / 0.54 | 0.3 / 0.54 | 0.2 / 0.36 | 0.2 / 0.36 |
| at 70℃ | 0.6 / 1.08 | 0.3 / 0.54 | 0.2 / 0.36 | 0.2 / 0.36 |
| at 100℃ | 0.3 / 0.54 | 0.3 / 0.54 | 0.3 / 0.54 | 0.2 / 0.36 |
| at 150℃ | 0.4 / 0.72 | 0.3 / 0.54 | 0.3 / 0.54 | 0.3 / 0.54 |
| at 180℃ | 0.3 / 0.54 | 0.4 / 0.72 | 0.4 / 0.72 | 0.5 / 0.9 |
| Gradient (±℃ / ±℉) 1) | ||||
| at -40℃ | 0.7 / 1.26 | 0.3 / 0.54 | 0.9 / 1.62 | 0.7 / 1.26 |
| at -25℃ | 0.6 / 1.08 | 0.3 / 0.54 | 0.9 / 1.62 | 0.7 / 1.26 |
| at -10℃ | 0.6 / 1.08 | 0.3 / 0.54 | 1.1 / 1.98 | 0.8 / 1.44 |
| at 25℃ | 0.8 / 1.44 | 0.4 / 0.72 | 0.7 / 1.26 | 0.6 / 1.08 |
| at 70℃ | 0.8 / 1.44 | 0.6 / 1.08 | 1.1 / 1.98 | 0.9 / 1.62 |
| at 100℃ | 0.9 / 1.62 | 0.7 / 1.26 | 1.01.2008 | 0.8 / 1.44 |
| at 150℃ | 1.1 / 1.98 | 1.2 / 2.16 | 0.8 / 1.44 | 0.9 / 1.62 |
| at 180℃ | 1.4 / 2.52 | 1.5 / 3.47 | 1.01.2008 | 1.3 / 3.43 |
| Heating time (min)2) | ||||
| from -40℃ to 25℃ | 30 | 31 | 20 | 21 |
| from -25℃ to 25℃ | 20 | 17 | 18 | 16 |
| from -10℃ to 25℃ | 17 | 13 | 13 | 15 |
| from 25℃ to 70℃ | 26 | 18 | 25 | 20 |
| from 25℃ to 100℃ | 28 | 21 | 26 | 25 |
| from 25℃ to 150℃ | 50 | 39 | 39 | 51 |
| from 25℃ to 180℃ | 56 | 63 | 46 | 56 |
| from -40℃ to 180℃ | 61 | 81 | 62 | 73 |
| from -25℃ to 120℃ | 52 | 39 | 47 | 47 |
| from -40℃ to 5℃ | 7 | 8 | 6 | 7 |
| from 5℃ to 125℃ | 44 | 59 | 40 | 49 |
| from 125℃ to 180℃ | 10 | 14 | 16 | 17 |
| Cooling time (min)2) | ||||
| from 25℃ to -40℃ | 37 | 29 | 28 | 46 |
| from 25℃ to -25℃ | 22 | 20 | 22 | 24 |
| from 25℃ to -25℃ | 22 | 20 | 22 | 24 |
| from 25℃ to -10℃ | 18 | 17 | 19 | 17 |
| from 70℃ to 25℃ | 20 | 22 | 23 | 23 |
| from 100℃ to 25℃ | 30 | 26 | 34 | 31 |
| from 150℃ to 25℃ | 95 | 92 | 110 | 160 |
| from 180℃ to 25℃ | 144 | 158 | 186 | 289 |
| from 180℃ to -40℃ | 172 | 198 | 192 | 297 |
| from 120℃ to -25℃ | 56 | 57 | 54 | 46 |
| from 180℃ to 125℃ | 110 | 133 | 144 | 228 |
| from 125℃ to 5℃ | 37 | 43 | 30 | 35 |
| from 5℃ to -40℃ | 25 | 22 | 18 | 34 |
| Recovery time after door was opened for 30sec. (min)2) | ||||
| at -40℃ | 6 | 6 | 6.5 | 9 |
| at -25℃ | 5 | 6.5 | 7 | 6.5 |
| at -10℃ | 6.5 | 6.5 | 8.5 | 6.5 |
| at 25℃ | 5.5 | 4.5 | 2 | 7 |
| at 70℃ | 8.5 | 8 | 9.5 | 7.5 |
| at 100℃ | 5.5 | 4.5 | 6 | 5.5 |
| at 150℃ | 6.5 | 7 | 9 | 3.5 |
| at 180℃ | 6.5 | 6.5 | 8 | 4 |
| Electrical data & Ordering information | ||||
| Electrical requirement (230V, 60Hz, 1Ph) | 26 | |||
| Electrical requirement (230V, 50Hz, 1Ph) | 26 | |||
| Electrical requirement (380V, 50/60Hz, 3Ph) | 12.5 | 19.6 | 19.6 | |