Jeio Tech AAH821332K TH3-ME-100 Constant Temp. and Humidity Chamber(100L) 230V, 50/60Hz
Jeio Tech AAH821332K TH3-ME-100 Constant Temp. and Humidity Chamber(100L)
Features
Optimized Model Configuration
- 3 series (KE, ME, PE) according to temperature / temperature and humidity control range.
TH3-KE; -40 ~ 100°C
TH3-ME; -20 ~ 100°C
TH3-PE; 0 ~ 100°C - Standard type Temp. & Humid chamber for selecting the optimized model according to test conditions.
- Dedicated stand for convenient use and moving.(option)
- Recorder for maintaining test records. (option)
Structural functional features
- Proven reproducibility and reliability with structural design optimized for miniaturization.
- Compact design for excellent space saving.
- Stable humidity control ability with applying differentiated external humidification system and convenient maintenance for cleaning testing space.
- Excellent thermal insulation and sealing to ensure stable temperature and humidity reproducibility even for repeated test use.
- The door part has a double sealing structure and excellent airtightness.
Use Convenience Features
- Convenient operation with 5.6 inch color touch display controller.
- Possible to check the operation status with graphs.
- Water level check function enables water replenishment at the right time.
- Water inlet is easy to use with easy cap.
- Efficient water supply and recirculation system minimizes water refilling even for long-term high-humidity tests.
- Operation and status monitoring from the front of the chamber.
- Easy observation of samples with tempered glass door with built in heating wire and LED lamp.
- By displaying the operation status of the device with LED colors, convenient to identify chambers status from a far.
- The cable port and silicone plug included as standard make it convenient to connect external equipment.
- Easy to clean as it is easy to remove refrigerator condenser grill, making it convenient to maintain efficiency of refrigeration.
- Integrated with 21 CFR Part 11 compliant LC DataKeeper, suitable for GMP audits (option)
Multifunction Control System
- Storage and documentation of programs.
- Save program information and back up the set values.
- RS-232 port (default) and RS-485 port (option) supported.
- Control and data processing by connecting up to 32 devices to PC at same time via RS-485 port.
- Reliable control with temperature/humidity auto-tuning and temperature calibration.
- PID zone subdivided into 9 zones for more precise control stability.
- Monitor and control via mobile app anytime, anywhere with LC Connected. (mobile monitoring system) (when purchased LC GreenBox)
- Convenient checking with useful graph display.
- Save data as excel and text files.
Outstanding Safety
- Electronic over temperature shutdown system.(advanced over temperature (A-OT) limiter)
Independent precision temperature sensor and controller for safer and more accurate over temperature shutdown. - General-type over power shutdown system. (B-OT, backup over temperature limiter)
A backup device for electronic system failure, mechanically preventing over temperature even when it occurs electronic errors. - Ensures that the surface temperature is kept below 60°C.
- Low and empty water level alarm and automatic shut off.
- Door open warning and automatic shut off.
- Overcurrent and short circuit protection of device.
- Fuse installed for each heater for safety.
- Auto stop when operating current of compressor is overloaded.
Specifications
| Model | TH3-ME-025 | TH3-ME-065 | TH3-ME-100 |
| Interior dimensions | |||
| Chamber volume (L / cu ft) | 25 / 0.9 | 65 / 2.3 | 100 / 3.5 |
| Width (mm / inch) | 350 / 13.8 | 400 / 15.7 | 500 / 19.7 |
| Depth (mm / inch) | 220 / 8.7 | 330 / 13.0 | 380 / 15.0 |
| Height (mm / inch) | 350 / 13.8 | 495 / 19.5 | 530 / 20.9 |
| Quantity of shelves (standard/max.) | 2/7 | 2/10 | 2/10 |
| Distance of between shelves (mm / inch) | 20 / 0.8 | 20 / 0.8 | 20 / 0.8 |
| Max. load per shelf (kg /lbs) | 25 / 55.1 | 25 / 55.1 | 25 / 55.1 |
| Permitted total load (kg / lbs) | 65 / 143.3 | 65 / 143.3 | 65 / 143.3 |
| Cable port (Ø, mm / inch) | 50 / 2.0 | 50 / 2.0 | 50 / 2.0 |
| Water tank capacity (L / cu ft) | 8 / 0.3 | 12 / 0.4 | 12 / 0.4 |
| Weight (kg / lbs) | 80 / 176.4 | 110 / 242.5 | 130 / 286.6 |
| Temperature data | |||
| Range without humidity (℃ / ℉) | -20 to 100 | -20 to 100 | -20 to 100 |
| / -4 to 212 | / -4 to 212 | / -4 to 212 | |
| Fluctuation (±℃ / ℉) 1) | 0.3 / 0.54 | 0.3 / 0.54 | 0.3 / 0.54 |
| Variation (±℃ / ℉) 1) | 0.5 / 0.9 | 0.5 / 0.9 | 0.5 / 0.9 |
| Heating time (0℃ to 90℃) (min) | 50 | 50 | 50 |
| Cooling time (90℃ to 0℃) (min) | 55 | 55 | 55 |
| Humidity data (※ Ref.: Temp./ Humid. control range graph) | |||
| Range (%RH) | 20 to 95 | 20 to 95 | 20 to 95 |
| Fluctuation (±%RH) 1) | 1 | 1 | 1 |
| Variation (±%RH) 1) | 3 | 3 | 3 |
| Electrical data & Ordering information | |||
| Electrical requirements (AC230V, 1P, 50/60Hz, A) | 7.7A | 9.6A | 13.8A |